NAC457.420. Tests of film processing equipment; action on results.  


Latest version.
  •      1. A test of film processing equipment used for mammography must be performed pursuant to 21 C.F.R. § 900.12 for each day that the equipment is in operation and before any clinical films are processed.

         2. The results of the tests performed pursuant to subsection 1 must be recorded, plotted on a control chart, evaluated and acted upon immediately after the test is completed and before any clinical films are processed. If the results are not within the applicable control limits, corrective action must be completed and verified as successful before any clinical films are processed.

     (Added to NAC by Bd. of Health, eff. 1-24-92; A by R033-06, 6-28-2006)