NAC459.568. Fluoroscopic X-ray systems: Activation of the tube.  


Latest version.
  • X-ray production in the fluoroscopic mode must be controlled by a device which requires continuous pressure by the fluoroscopist for the entire time of any exposure. When recording serial fluoroscopic images, the fluoroscopist must be able to terminate the X-ray exposure at any time, but means may be provided to permit completion of any single exposure of the series in process.

     [Bd. of Health, Radiation Control Reg. § 6.5.2, eff. 2-28-80]